Abstract
The presence of cuff imbalance degrades the signal-to-interference (ENG/EMG) ratio in tripolar nerve cuff electrode recordings. Known causes of cuff imbalance include inhomogeneous tissue growth after cuff implantation and cuff manufacturing tolerances. In this paper, we report on an additional contribution to cuff imbalance that stems from variations in orientation and distance of the tripolar cuff relative to the external interference source. The latter is represented here by a dipole. Interference amplitude is also shown to depend on orientation and distance variations, here both factors included in the term "proximity." The study was conducted using field simulations and saline-bath experiments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.