Abstract

The Sieverts technique is a popular method of measuring the uptake of gas on a sample and is an important tool for the characterisation of hydrogen uptake by potential hydrogen storage materials. An analysis of the consequences of errors in the calibration of volumes in this technique has been made using hydrogen absorption on an AB2 alloy as an example. Trends in the error in the uptake as a function of errors in the volumes have been explained in terms of the equations for uptake. Significant advantages are found for calibration techniques for which the ratio of volumes is an intrinsic measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.