Abstract

We investigated whether the contamination of samples with glucose subsequently tested for haemostasis affected the results, including prothrombin time, activated partial thromboplastin time and fibrinogen concentration. Venous blood was collected from 12 healthy subjects and divided into four aliquots, which were subjected to different degrees of contamination with standard glucose solution (0%, 5%, 10%, 20%). With increasing glucose contamination, prothrombin time increased from mean (SD) 11.0 (0.7) s to 11.2 (0.7) s, 11.5 (0.7) s and 12.2 (0.8) s, all p < 0.001. Activated partial thromboplastin time decreased from 32.3 (0.9) s to 30.9 (0.8) s, 30.8 (0.8) s, and 29.7 (0.7) s, all p < 0.001. Fibrinogen concentration decreased from 3.8 (0.7) g.l(-1) to 3.7 (0.6) g.l(-1), 3.6 (0.6) g.l(-1), and 3.4 (0.6) g.l(-1), all p < 0.001. Bias was clinically meaningful from 5% contamination for activated partial thromboplastin time, 10% contamination for prothrombin time and 20% contamination for fibrinogen concentration. We conclude that if glucose contamination of haemostasis samples is suspected or has occurred, the specimens should not be analysed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.