Abstract

Blood viscosity during operation of ventricular assist device (VAD) can be changed by various conditions such as anemia. It is known generally that the blood viscosity can affect vascular resistance and lead to change of blood flow. In this study, the effect of fluid viscosity variation on hemodynamic energy was evaluated with a pulsatile blood pump in a mock system. Six solutions were used for experiments, which were composed of water and glycerin and had different viscosities of 2, 2.5, 3, 3.5, 4, and 4.5 cP. The hemodynamic energy at the outlet cannula was measured. Experimental results showed that mean pressure was increased in accordance with the viscosity increase. When the viscosity increased, the mean pressure was also increased. However, the flow was decreased according to the viscosity increase. Energy equivalent pressure value was increased according to the viscosity-induced pressure rise; however, surplus hemodynamic energy value did not show any apparent changing trend. The hemodynamic energy made by the pulsatile VAD was affected by the viscosity of the circulating fluid.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.