Abstract

Background and aims Determining the proper length of the root canals is essential for successful endodontic treatment. The purpose of this in vitro study was to evaluate the effect of file size on the accuracy of the Raypex 5 electronic apex locator for working length determination of uninstrumented canals. Materials and methods Twenty maxillary central incisors with single straight canals were used. Following access cavity preparation, electronic working length by means of Raypex 5 apex locator and actual working length were determined. Data were analyzed using ANOVA with repeated measurements and LSD test. Results There was no significant difference between electronic and actual working lengths when a size 15 K-file was used. Conclusion Under the conditions of the present study, a size 15 K-file is a more suitable size for de-termining working length.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.