Abstract
RBC transfusions are often administered using externally applied pneumatic pressure. The effect of this practice on RBC hemolysis was studied during simulated transfusion performed in the laboratory. RBC transfusions were performed sequentially via 16-, 18-, 20-, and 22-gauge cannulas with the application of a pneumatic pressure device at pressures of 150 and 300 mm Hg. Hb concentration, Hct, RBC count, free Hb, potassium, and LDH concentrations were measured in 5 mL of transfusate. Forty-seven RBC units, 20 fresh units (mean age, 10.8 days) and 27 old units (mean age, 28.9 days) were studied under all conditions. Multiple regression analysis revealed that the age of the unit and the external pressure applied were significant determinants of the degree of hemolysis. The application of an external pressure device results in minimal destruction of transfused RBCs, even under the most stringent conditions examined. Thus, external pressure application to expedite an RBC transfusion is likely to be a safe procedure for the majority of patients. Unusual clinical situations, such as massive transfusion in pediatric patients, should be specifically examined to confirm the safety of this procedure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.