Abstract

The influence of electrode compression by the cervix on transcutaneous CO2 measurements (tc-Pco2) was studied during 20 deliveries by simultaneous measurements with two electrodes, one positioned under the cervical edge and the other free of the cervix. At a cervical dilation of less than 8 cm, significantly higher values were measured by the electrode exposed to cervical compression, and tc-Pco2 fluctuations with uterine contractions were observed. This study confirms that the tonsure effect usually does not interfere with tc-Pco2 monitoring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.