Abstract
Purpose: To evaluate the effect of dose reduction on image quality in digital radiography using a flat-panel detector. Materials and Methods: Digital radiographs of 30 rabbits were obtained at two different dose levels (33.23 Gy for the standard dose group and 20.09 Gy for the reduced dose group). The amorphous selenium-based flatpanel detector system had a panel size of 78.5 inches, a matrix of 12801536 (pixels?), and a pixel pitch of 138 m. Four observers evaluated the soft-copy images on a high-resolution video monitor (256020488 bits) in random order. The observers rated the visibility of 13 different anatomic structures on a 5-point scale, viz. the retrocardiac lung, subdiaphragmatic lung, heart border, diaphragmatic border, proximal airway, unobscured lung, liver border, kidney border, bowel gas, flank stripe, ribs, and vertebrae in the mediastinal and abdominal regions. Statistical significance was determined using Wilcoxon’s signed rank test. Results: There was no statistically significant difference in the visibility of the anatomic structures on digital radiography between the standard and reduced dose groups. Conclusion: Digital radiography using an amorphous selenium-based flat-panel detector can preserve the image quality, even though the does is reduced to 40% of the standard level.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.