Abstract

Objective To evaluate the effect of two kind oxygenators on polymorphonuclear neutrophil (PMN) apoptosis, and to give a method of inhibting the systemic inflammatory response syndrome(SIRS) to cardiopulmonary bypass(CPB). Methods Sixty VSD patients undergoing open heart surgery with CPB were randomly divided into two groups(n = 30): group B(bubble oxygenator), group M(membrane oxygenator). Blood samples were drawn from the central venous line before starting CPB, at the end of CPB, 4, 8 and 24 h after CPB. The PMN counts were performed by blood cell counter. PMN apoptosis and the expression of CD11b were evaluated by flow eytometry. The plasma concentration of elastase was determined by enzyme linked immunosorbent assays. Results The rates of PMN apoptosis were significantly reduced (P<0.05) and the PMN counts, the expression of CD11b and the plasma con-centration of elastase were remarkably increased after CPB(P<0.05), but at the end of CPB, 4 h and 8 h after CPB the rate of PMN apoptosis was higher in group M than that in group B. PMN counts, the expression of CDI lb and the plasma concentration of elastase were higher in group B than in group M (P<0.05). The PMN counts were higher in group B than in group M 24 h after CPB (P< 0.05). Conclusion XiJian-Ⅱ membrane oxygenator can increase PMN apoptosis and depress systemic inflammatory response better than XIJing-90 bubble oxygenator. Key words: Cardiopulmonary bypass; Inflammatory response; Membrane oxygenators; Bubble oxygenators