Abstract

Variability in the magnitude of the impaction force applied by surgeons to assemble a hip prosthetic head to a femoral stem could be a cause of increased wear in taper junctions. This study investigates the effect of varying the magnitude of the assembly force on fretting wear at the taper over a 10 year period using a 3D finite element model and wear algorithm. It is demonstrated that an increase in assembly force results in a reduction in fretting wear and it is recommended that surgeons should apply an impact force of at least 4kN to minimise wear rates. The wear patterns and wear rates presented are comparable with observation and measurement of those seen in retrieved prostheses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.