Abstract

The effect of cycle length upon refractory period of the atria was measured in 34 patients. The effect of abrupt change in paced cycle length on atrial refractory period was measured by introducing after every eighth paced cycle (CLP) a test cycle (CLT) at an interval which was either 100 ms longer (CLT = CLP + 100 ms) or shorter (CLT = CLP - 100 ms) than the paced cycle. These findings confirm that: (1) atrial refractory periods shorten with decreasing cycle length during steady state pacing; (2) an abrupt change in paced cycle length results in a change in refractory period. This change is similar in direction but smaller in magnitude than the change which occurs in response to steady state pacing; and (3) there is no correlation between intraatrial conduction time and refractoriness.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.