Abstract

The width of the blood pressure cuff was found to have an effect on the accuracy of systolic blood pressure measured in the arm and thigh in neonates. Data were collected in 26 infants of a variety of birth weights and gestational ages. They show that the effect of cuff size can be minimized by selecting a cuff width equal to approximately 50% of the extremity circumference.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.