Abstract

The effect of column end structure and bed heterogeneity of six commercially available reversed-phase chromatographic columns for fast liquid chromatography with different column packing materials – such as fully porous (Waters XBridge C18 with 1.7 μm particles) and superficially porous (Waters CORTECS C18 with 1.6 μm particles), with column dimension of 2.1 × 50, 100 or 150 mm were tested with flow-reversal method. The method includes arresting the flow when a non-retained marker (thiourea) has penetrated to a given distance into the column and then reversing the column. Hence, when the flow has been restarted, the sample is eluted at the same end of the column where it entered. The experiments showed that all columns are axially heterogeneous, and some differences could be observed between the two respective column ends. Furthermore, we can conclude that the shorter columns are axially more homogeneous than the longer ones, thus the column length is an influencing factor on the column packing procedure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.