Abstract
The precise determination of the dead time is essential in a chromatography system as it is a primary parameter for the determination of other secondary parameters such as the adjusted retention time, relative retention time, retention factor and retention index. Several of the indirect methods used for the determination of the dead time in this study were iteration, nonlinear, spreadsheet and statistics methods, which were implemented using the ANSI C programming language. The calculation of each method was tested with temperature and column variations for measuring the retention time of a n-alkane homologous series and accuracy analysis of each mathematical method (indirect method) to the marker substance (direct method). Changes in the temperature and column variations (type, polarity and column length) affected the calculation of the dead time values but did not affect its accuracy. The value of the dead time generated by the non-linear method was relatively high, with errors above 10%, while the other methods utilized are quite good with errors below 8% regardless of the column and temperature variations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.