Abstract

This in vitro study examined the effect of circuit compliance on delivered ventilation (VE) using a time-cycled, volume controlled circle system in an infant lung model. A Bio-Tek ventilator tester set to simulate normal and abnormal lung compliance measured VE delivered by the Narkomed 2B system. Circle circuits of varied compliance (2.75, 1.22 and 0.73 microliters.cm H2O-1) were tested. Tidal volume was adjusted to peak inflation pressures (PIP) of 20, 30, 40, and 50 cm H2O with three circuits, two lung compliances, and four different size tracheal tubes (TT) (2.5, 3.5, 4, 4.5 mm ID). Data were analysed using the multiple regression technique. Delivered VE was directly related to PIP and lung compliance. Delivered VE was not affected by the choice of circuit. TT size had minimal effects on VE when lung compliance was low; TT size was a more important factor when test lung compliance was normal. Extrapolating this data to the clinical setting, adequate ventilation of infants can be achieved with an adult circle system if an appropriate PIP is chosen, regardless of the compliance of the circuit used. Infants with poor lung compliance may require very high PIP for adequate ventilation.