Abstract

Surface resistivity is an important parameter of multi-electrode CZT detectors such as coplanar-grid, strip, or pixel detectors. Low surface resistivity results in a high leakage current and affects the charge collection efficiency in the areas near contacts. Thus, it is always desirable to have the surface resistivity of the detector as high as possible. In the past the most significant efforts were concentrated to develop passivation techniques for CZT detectors. However, as we found, the field-effect caused by a bias applied on the cathode can significantly reduce the surface resistivity even though the detector surface was carefully passivated. In this paper we illustrate that the field-effect is a common feature of the CZT multi-electrode detectors, and discuss how to take advantage of this effect to improve the surface resistivity of CZT detectors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.