Abstract

AbstractThe concentration sensitivity of a thermal conductivity detector (TCD) depends, among other factors, on the amount of sample mixture in the detector's sensing cell. Since the cell volume has to be appropriately matched with column diameter, it makes the concentration sensitivity of a TCD dependent on column diameter and, therefore, on the speed of gas chromatography. Through reduction of column diameter, higher speed tends to lead to a reduction in the concentration sensitivity of the cell. The factor which the most directly affects the concentration sensitivity of a TCD cell is the heat power conducted through the cell. The higher the power, the greater the sensitivity. The limit of detection of a TCD depends on the concentration‐sensitivity of its cell and on the level of statistical errors in the measurement. The errors increase with increasing analysis speed. As the column diameter is reduced, the errors cause additional worsening (on top of the decrease in concentration sensitivity) of the detection limit, dynamic range, and other performance characteristics of the TCD.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.