Abstract
Noninvasive in vivo detection of gaseous microemboli in the middle cerebral artery, by transcranial Doppler ultrasound, was used to determine the effect of filtration in the arterial catheter using 25- and 40-μm filters and bubble oxygenators in patients undergoing cardiopulmonary bypass surgery. Eighteen patients undergoing coronary artery bypass surgery were studied using a closed cardiac (unvented heart) model. Group 1 patients (no filters) had the highest incidence of gaseous microemboli, as indicated by the ultrasound microemboli index, at both high and low oxygen flow rates. Group 2 patients (40-μm filters) had a significantly lower microemboli index, particularly at low oxygen flow rates ( t = 4.9, p < 0.001). The 25-μm group patients had the lowest values of all. No microemboli were detected at low oxygen flow rates, and microemboli were detected in only 0.1% of the samples at high oxygen flow rates. Additionally, observations on vented hearts in 3 patients undergoing cardiac valve surgery indicate that the origin of gaseous microemboli may be air trapped inside the heart.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.