Abstract
Forearm crutches are a commonly used assistive device to aid the ambulation of people with lower limb impairment. Due to the structure of the joints of the upper limb, the loading patterns during crutch-walking can be detrimental to the user. Shock-absorbed or compliant crutches may reduce ground contact impact, potentially benefitting the user. In this study, the authors performed a biomechanical evaluation of a shock absorber crutch, which uses an elastomeric system to absorb impact. Ground reaction forces (GRFs) and spatiotemporal parameters were compared between shock absorbed and standard crutches. Results indicate no significant differences to spatiotemporal parameters and peak ground reaction forces between the crutches. There is an initial impact spike during crutch strike when using shock absorbed crutches, after which the rate of force development is moderately decreased when compared with a standard crutch. The practical benefit of a small reduction in force is questionable and more work must be undertaken to further optimise shock absorber design.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.