Abstract

This study aims to evaluate the effect of pre-reconstruction process for low tube voltage computed tomography (CT) on image quality of filtered backprojection (FBP) reconstruction. Small and large quality assurance water phantoms (19 and 33cm diameter) were scanned on a third-generation dual-source CT with 70 kVp and 120 kVp at various dose levels. Image quality was assessed in terms of the noise power spectrum (NPS) and task-based transfer function (TTF). NPSs and TTFs in the small phantom were comparable between 70 and 120 kVp protocols. In the large phantom, the curves of the NPS changed and the TTF decreased even at the high-dose levels for 70 kVp protocol compared to 120 kVp protocol. Our results indicated that the pre-reconstruction process is performed in low tube voltage CT for large objects even for the FBP reconstruction and has an effect on the image quality.