Abstract

AbstractObjective:To study the effect of a portable HEPA-filtered air exhaust system (Stackhouse Freedom Surgical Helmet System) on airborne microbial contamination in a modern conventional operating room.Design and Setting:Microbial air sampling was done with a two-stage Anderson sampler at the wound site during 46 total joint replacements. All operations were performed by the same surgeon in the same operating room at a large community hospital.Results:In 18 cases done without air exhaust hoods, the number of bacterial and fungal colony-forming units (CFU) ranged from 0.6 to 11.7 (mean, 3.6). Air sampling during 28 operations with the operating team in air exhaust hoods revealed a mean of 3.6 CFU (range, 0 to 11.4). Bacterial CFU averaged 3.4 without hoods and 3.2 with exhaust hoods. Coagulase-negative staphylococci were the most common isolates (48% of isolates with hood, 55% without hood). No infections occurred.Conclusion:We conclude that these air exhaust hoods did not lower airborne microbial contamination detectable with this air sampling method, as compared to standard head cover and mask, in a modern conventional operating room.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.