Abstract
For personnel inexperienced in airway management, supraglottic airway devices may be the first choice in an emergency. Changing head position is known to reduce the seal pressure of a laryngeal mask airway. The aim of this study was to investigate whether the use of a cervical collar improves the stability of airways secured with the LMA Supreme™ (The Laryngeal Mask Company Limited, Mahé, Seychelles). In this crossover trial, the primary endpoint was the difference in the seal pressure of the LMA Supreme in anaesthetised patients in maximum passive extension of the neck, with and without a cervical collar. The median (IQR [range]) seal pressure was 18 (13.8-22.1 [0-30]) cmH(2) O in maximum passive extension without a cervical collar. With a cervical collar in place, the seal pressure increased to 28 (22.8-30 [17-30]) cmH(2) O (p<0.001). In the neutral head position, the seal pressure was 22 (17.6-24.5 [12-30]) cmH(2) O without and 27 (22-30 [12-30]) cmH(2) O with a cervical collar in place (p<0.001). We found that a cervical collar stabilises the airway with an LMA Supreme in place and we recommend this combination for (pre-hospital) emergency cases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.