Abstract

Pupil diameter measurement is crucial for physical assessment and disease monitoring in a health and nursing care situation. A general penlights (GPLs) is frequently used and allow for an approximate and indirect measurement of the pupil diameter. Health caregivers or nurses generally have less confidence in the value of the pupil diameter measured using the GPL. The Advanced Penlight (APL) is a new device designed for accurate measurement of the pupil diameter. The purpose of the presented research was to compare the accuracies and operational times of the pupil diameter measurements by means of the GPL and APL. One-group post-test and single-blind study designed was used in this study. The innovation of the APL is the addition of a perspective measurement ruler (PMR) attached to one side of the penlight that allows precise measurement of the pupil diameter before and after pupillary contraction. The PMR can be rotated by any angle for adaptation to the measurement conditions. After standard pupil diameter measurements by a refractometer (RM) were performed on a subjects, ninety study participants measured the pupil diameters of the same subject separately by the GPL and APL. A self-administered questionnaire was used to assess the opinions of the participants after using the GPL compare to the APL. The mean age of the participants was 20.01 (SD = 0.47) years and 83% of them were female senior nursing students. There were no statistically significant differences between the average values of pupil diameters measured by the APL and the RM. Compared to the GPL, the pupil diameter measured by APL was much similar to the RM measurement. The average operational time was 8.72 seconds shorter (t = -3.81, p = 0.001) for the APL measurement compared to the GPL measurement. The average scores of convenience and confidence on pupil diameter measurements of questionnaire were higher for the APL compared to the GPL. The APL can increase the accuracy and save operating time of pupil diameter measurement and thereby promote the quality of health assessment and nursing care practice.

Highlights

  • Pupil size and reactivity assessment is a regular health and nursing care in general ward, emergency room (ER), and intensive care unit (ICU) [1,2]

  • The results indicated the pupil diameters measured by general penlights (GPLs) and the RM were different

  • The electronic and automated pupillometry is a more reliable instrument in patients’ pupil assessment, the GPL is still frequently used in clinical care situations to measure the pupil diameter and pupil reflex of Operational time

Read more

Summary

Objectives

The purpose of this study is to compare the accuracy and the operational time of pupil diameter measurement by the GPL and a new design penlight

Methods
Results
Discussion
Conclusion