Abstract

A pink plastic packaging bag sample from a enterprise was preliminarily screened via X-ray fluorescence spectroscopy method (XRF), it was found that the lead content exceeded the limit of relevant national standards. Subsequently, the total lead content in the packaging bag was determined via atomic absorption spectrophotometry pretreated by dry ashing method (DA-AAS) and inductively coupled plasma–mass spectrometry with microwave digestion method (MD-ICP-MS), respectively. The working curve equations were obtained by linear least squares method with the calculated linear correlation coefficients r above 0.9998 and the relative standard deviation (RSD) of the determination results from 1.0% to 6.2%, which was in good agreement with the result of XRF.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call