Abstract

We report our technique for hemibody skin electron irradiation with a custom‐made plywood shield. The technique is similar to our clinical total skin electron irradiation (TSEI), performed with a six‐pair dual field (Stanford technique) at an extended source‐to‐skin distance (SSD) of 377 cm, with the addition of a plywood shield placed at 50 cm from the patient. The shield is made of three layers of standard 5/8” thick plywood (total thickness of 4.75 cm) that are clamped securely on an adjustable‐height stand. Gafchromic EBT3 films were used in assessing the shield's transmission factor and the extent of the dose penumbra region for two different shield‐phantom gaps. The shield transmission factor was found to be about 10%. The width of the penumbra (80%‐to‐20% dose falloff) was measured to be 12 cm for a 50 cm shield‐phantom gap, and reduced slightly to 10 cm for a 35 cm shield‐phantom gap. In vivo dosimetry of a real case confirmed the expected shielded area dose.PACS number(s): 87.53.Bn

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.