Abstract

BackgroundNasal intubation using a fiberoptic scope is a useful technique. In clinical practice, we have experienced difficulty in advancing the endotracheal tube (ETT) over the fiberoptic scope because of resistance to the passage of the ETT against rotation in the nasal cavity, when it gets hung up on structures of the laryngeal inlet. Several maneuvers have been proposed to overcome this difficulty. The gap between the tip of the ETT and the fiberoptic scope can be reduced using a thicker fiberoptic scope and a thinner ETT. Moreover, simultaneous rotation of the fiberoptic scope and ETT could lead to successful intubation by reducing impingement on the ETT. However, the discrepancy between these rotation angles is unclear. This observational prospective study aimed to investigate the discrepancy in the rotation angle between the ETT and fiberoptic scope during nasal intubation.MethodsThe patients (aged 20–80 years) who underwent nasal intubation for oral and maxillofacial surgery participated in three sizes of preformed nasal ETT and were intubated using a fiberoptic scope. They were divided into three groups; the ETT internal diameter (ID) 6.5 mm (6.5 group), ID 7.0 mm (7.0 group), and ID 7.5 mm (7.5 group). The ETT was then inserted through the nasal cavity into the pharynx. After the fiberoptic scope was advanced through the ETT above the glottis, simultaneous rotation by both the proximal end of the fiberoptic scope and ETT was performed in 90° and 180° in both right (clockwise) and left (counterclockwise) directions, and the rotation angle at the distal end of the ETT was monitored using a video laryngoscope (Pentax-AWS).ResultsA total of 39 patients were included in the study. When both the proximal end of the fiberscope and ETT were simultaneously rotated by 90°, in the 6.5 group (n = 13), the distal end of the ETT rotated by 47.8 ± 1.5°. In the 7.0 °group (n = 13), the distal end of the ETT rotated by 45.5 ± 1.0°. In the 7.5 group (n = 13), the distal end of the ETT rotated by 39.9 ± 1.0°. When the proximal end of the fiberscope and ETT were rotated by 180°, in the 6.5 group, the distal end of the ETT rotated by 166.2 ± 2.5°. In the 7.0 group, the distal end of the ETT rotated by 145.7 ± 2.2°. In the 7.5 group, the distal end of the ETT rotated by 115.1 ± 2.0°. All rotation angles in the distal end of the ETT were significantly lower than those in both the proximal end of the fiberscope and ETT (p < 0.05). Rotating right by 180° was significantly different among the three groups (p < 0.05), although rotating right by 90° was not significantly different. Similar results were obtained for the left rotation.ConclusionSimultaneous rotation by the proximal end of the ETT and fiberscope above the glottis for the nasal approach induced significant differences in the distal end of the ETT. The larger tube lagged by the resistance of the nasal passages during rotation. Therefore, the ETT does not rotate as much as the rotation angle.Trial registrationThis prospective observational study was conducted after receiving approval from the Ethics Review Board of Kyushu University Hospital (Approval No. 30–447).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call