Abstract

The diagnosis of right ventricular perforation by an endocardial pacemaker electrode should be suspected when failure of pacing occurs without electrode displacement. Although a number of changes occur on the standard electrocardiogram (ECG), none of these are diagnostic. The intracardiac electrogram performed during electrode withdrawal is not only diagnostic of perforation but can also aid in electrode positioning. Two case reports highlight these changes in the intracardiac electrogram. The first case also illustrates that, with electrode perforation, the ability to sense the intrinsic intracardiac electrical activity may be retained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.