Abstract

An ultra performance liquid chromatography–atmospheric pressure chemical ionization–mass spectrometry (UPLC–APCI–MS) method was developed for the separation and quantification of adrenal steroid metabolites from heterologous expression media. Steroids were extracted by liquid–liquid extraction, separated on a Waters UPLC BEH C18 column, ionized by APCI, and detected using a triple quadrupole mass spectrometer in APCI positive mode with single ion monitoring. The incorporation of UPLC enabled the detection of seven structurally closely related steroids at between 5 and 40 ng/ml using run times of 11 min. The adrenal steroidogenic enzyme cytochrome P450 17-hydroxylase/17,20-lyase (CYP17) was expressed in the yeast Pichia pastoris and in nonsteroidogenic COS-1 cells, and used as a model system to evaluate the detection and quantification of adrenal steroid metabolites by UPLC–APCI–MS.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.