Abstract

The performances of new "total" aerosol samplers for use in workplaces are required to match the inhalability criteria as contained in the latest recommendations of the International Standards Organization (ISO) and the American Conference of Governmental Industrial Hygienists (ACGIH). In the past, practical evaluations have been carried out under idealized conditions in wind tunnels, and there is now the need to extend these to more realistic workplace conditions. This paper describes a new test system that was designed and built for this purpose. It consisted of a life-size mannequin mounted on a trolley so that it can be taken to and wheeled around in workplaces. The mannequin itself incorporated a robotic arm so that, under joystick control, it can be made to simulate a range of worker movements, orientations, and attitudes. An electronically controlled, compact breathing machine provided a range of typical breathing parameters for the mannequin. The pump also provided air movement for a number of personal samplers that were mounted on the torso of the mannequin and tested in that position. Sampler performance should be assessed by comparing directly the aerosol collected by the sampler with that inhaled by the mannequin (and collected on filters inside the head).