Abstract

A new parallel liquid chromatography/tandem mass spectrometry (LC/MS/MS) system has been developed, in which the mass detector was shared between two staggered parallel chromatographic runs. Since the chromatography for biofluids assay generally requires good analyte retention and thus tends to leave large blank chromatographic windows, this parallel system allowed the efficient use of the mass detector during these blank windows, resulting in significantly improved sample throughput. Also, in order to remove the bottleneck in sample extraction for this parallel separation system, a high-flow extraction device was used to perform on-line extraction. This allowed for the direct injection of biofluids onto the system. The performance and capability of this system was evaluated in tests that contained a single analyte (oxazepam) and multiple analytes (12-in-1). The results indicated that the data generated from this system were comparable to those obtained on a conventional single-column system. An application of the system for high-throughput pharmacokinetic screening of drug candidates was also demonstrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.