Abstract

Mycoplasma are small bacteria that have the ability to penetrate sterilizing-grade filters. Filtration of large-scale mammalian and bacterial cell culture media is an example of an industry process where effective filtration of mycoplasma is required. The Parenteral Drug Association recognized the absence of industry standard test parameters for evaluating mycoplasma clearance filters by filter manufacturers and formed a task force to formulate such a consensus among manufacturers. The use of standardized test parameters by filter manufacturers, including the preparation of the culture broth, will facilitate the end user's evaluation of the mycoplasma clearance claims provided by filter vendors. However, it is still important to recognize filter performance will depend on the actual conditions of use; therefore end users should consider, using a risk-based approach, whether process-specific evaluation of filter performance may be warranted for their application.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.