Abstract

The development, verification, and comparison study between LC–MS libraries for two manufacturers’ instruments and a verified protocol are discussed. Compounds in the libraries are among those considered by the U.S. EPA Office of Water as threats to drinking water including pesticides, drugs of abuse, and pharmaceuticals. The LC–MS library protocol was verified through an inter-laboratory study that involved Federal, State, and private laboratories. The results demonstrated that the libraries are transferable between the same manufacturer's product line, and have applicability between manufacturers. Although ion abundance ratios within mass spectra were shown to be different between the manufacturers’ instruments, the NIST search engine match probability was at 96% or greater for 64 out of 67 compounds evaluated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.