Abstract
A new EELS (electron energy loss spectroscopy) real-time elemental mapping system has been developed for a dedicated scanning transmission electron microscope (STEM). The previous two-window-based jump-ratio system has been improved by a three-window-based system. It is shown here that the three-window imaging method has less artificial intensity in elemental maps than the two-window-based method. Using the new three-window system, the dependence of spatial resolution on the energy window width was studied experimentally and also compared with TEM-based EELS. Here it is shown experimentally that the spatial resolution of STEM-based EELS is independent of the energy window width in a range from 10 eV to 60 eV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.