Abstract

Abstract Trace level Volatile Organic Compounds (VOCs) in ambient air are normally determined according to EPA Method TO-14. This method describes the analysis in ambient air of 41 VOCs, ranging in boiling point from -29 to 215°C. It covers a concentration range from 0.2 to 20 parts per billion, volume/volume (ppb), and specifies sample enrichment of a 400 mL air sample on glass beads at -160°C. While this sample volume provides sub-ppb levels of VOC detection for target analytes when using a quadrupole mass spectrometer detector in SIM mode or when using GC detectors, the identification of non-target analytes may only be done in full scan mode for higher concentrations. Also with this sample volume a Nafion dryer is needed for water removal thereby lowering the recovery of polar VOCs. Because of the very high sensitivity of the ion trap MS, relatively small air volumes (60 mL) are adequate to obtain the required or lower detection levels. An integrated air analysis system based on a GC-ion-trap MS has be...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.