Abstract
Scatter functions and relative depth-dose are computed for a variety of X-ray spectra and exposure geometries of dental interest. The method takes into account the attenuation produced by the lead-foil backing in intraoral film packets and makes use of an analytical model to compute the zero-area depth-dose based on known spectra and attenuation coefficients.The resulting data show that the relative amount of scatter at the film plane for a fixed exposure at the tissue surface is significant and approximates half the dose produced by the primary beam neasured at the same depth. The foil backing reduces scattered radiation reaching the film plane by an amount which is significantly influenced by the energy spectrum of the primary beam.Signal-to-noise ratios per unit area estimated from the experimental data for non-screen film show that the parameters controlled in this investigation have a relatively small but statistically significant effect on the information carrying capacity of the radiographic system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The Journal of Nihon University School of Dentistry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.