Abstract

A non-destructive analytical method for oxygen in silicon by both helium-3 and α-particle activation analysis was applied to different samples, containing 0.05–10 p.p.m. of oxygen. The accuracy and precision of the method were studied. For concentrations above 1 p.p.m., the average coefficient of variation is about 6% for α-activation and 13% for helium-3 activation. For concentrations of about 60 p.p.b., the corresponding figure is about 20%. The results of both methods, which agree within 8%, are compared to infrared analysis. The determination limit of the method is 4 p.p.b. for helium-3 activation and 22 p.p.b. for α-activation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.