Abstract

The paper deals with the methodology, quantification and comparison of measurement deviations for two types of the ceramic shard. The ceramic material has better polarization properties than concrete material. For impedance measurement the ceramic shard is more appropriate. The IS method diagnostics based on processing of measured data which aims evaluating the absolute error and the relative error (deviation) of the loss factor values. Results show the spectrum parts which are useful in the lower frequency range only. When Agilent 33220A and Agilent 54645A are used to carry out the IS, the most important factor from the viewpoint of the tgδ(f) determination accuracy is the accuracy of the phase shift between the excitation signal and that at the measured system output. Other factors are the measurement accuracy of the electric capacity and resistance of the oscilloscope input channel.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.