Abstract

A critical comparison of the application of differential pulse voltammetry and anodic stripping voltammetry to the determination of micro amounts of copper in silicon is described. The anodic stripping technique offers advantages when a dropping mercury capillary with a long drop time is used. The method recommended allows the determination of copper in silicon with a precision of ± 5 %; the limit of determination is about 1μg g -1. Calibration graphs are linear in the range 0–0.2 μg Cu ml -1. Methods for the dissolution of silicon are also compared.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.