Abstract
A simple method is described for the continuous determination of chemical oxygen demand (COD) in wastewater samples by flow injection analysis. Samples are injected into a water stream which merges with an acidic dichromate carrier solution. After reaction in a PTFE coil at 120°C, absorbances are measured at 445 nm. D-Glucose is satisfactory as a standard substance for COD in a variety of wastewaters. A sampling rate of 15 samples per hour can be achieved, and the detection limit and precision are 5 mg l -1 as COD and 0.4%, respectively. Chloride concentrations of ⩾ 100 mg l -1 interfere slightly unless silver and/or mercury salts are added. COD values for various wastewater samples correlate well with those obtained by standard methods using dichromate or permanganate.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.