Abstract
Abstract A simple flow injection analysis (FIA) method of albendazole (ALB) using UV detection and a HPLC method are described in this study. For the FIA method, the best solvent system was found to consist of methanol∶HCl∶water (49∶1∶50). The limit of detection (LOD) and limit of quantitation (LOQ) were calculated to be 1.2×10−8 and 3.7×10−8 mol · L−1 (n=8), respectively. For the HPLC method, the mobile phase is a mixture of methanol:acetate buffer (0.05 M, pH:5.8) (70∶30 v/v). The retention time obtained was 8.07 min. The limit of detection (LOD) and limit of quantitation (LOQ) were calculated to be 1.1×10−8 and 3.8×10−8 mol · L−1 (n=8), respectively. The proposed methods were applied to the determination of ALB in the suspensions. The results were compared with those obtained from UV spectrophotometry. The validation studies were realised by the related applications and the results were evaluated statistically.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Liquid Chromatography & Related Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.