Abstract

Abstract A flow-injection analysis method for the determination of a small amount of protein has been established by the use of the ultrasonic chemiluminescence of luminol. The present method was based on the phenomenon that the catalytic activity of cobalt(II) for the ultrasonic chemiluminescence of luminol decreased in the presence of protein. The apparatus consisted of only a luminol solution line equipped with an ultrasonic irradiation bath and a cobalt(II) catalyst solution line equipped with a sample inlet. Under optimum conditions, similar calibration curves were obtained for bovine serum albumin, bovine serum γ-globulin, human serum albumin, and human serum γ-globulin. According to the present method, these proteins could be determined in the concentration range 5.0×10−6–1.0×10−1 g dm−3, with a detection limit of 200 pg (injected sample volume 40 mm3, S/N=2) and a coefficient of variation of 2.9% (n=8) at a rate of about 60 samples per hour.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.