Abstract

The possibility of localisation of an unknown neutron source in various bulky homogeneous media (box) was studied. Previous study showed that for the planar case, neutron source could be localised with accuracy better than 7%, by using the count rates ratio of two (3)He detectors, positioned on the opposite faces of the box. The influence of the deviation of the source from the virtual plane connecting the two (3)He detectors was studied. It was found that for detectors placed on the centres of the sample faces, larger distance of the source from the central line leads to larger error, up to 15%. It was found that while using the measurement of the ratio R(0) of external source placed on the box face coplanar with unknown source, accuracy does not depend on the distance from central line. Different ways to determine the plane of the unknown source were investigated. It was found that it is impossible to determine it by looking for maximum count rate of one detector or geometric mean of a pair of detectors. The experimental results were confirmed with Monte-Carlo N-particle (MCNP) simulations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.