Abstract

This paper introduces the design of permanent magnet that can been used to small nuclear magnetic resonance (NMR) system, and its static magnetic field simulation analysis uses Ansoft's Maxwell software. According to the theory of magnetic circuit design and the performance requirements of magnetic field, An H-style permanent magnetic actuator has been designed, and it can generate uniform magnetic field larger than 0.4 T in the interested region of this actuator. The static magnetic field simulation analysis of this permanent magnetic actuator has done by Ansoft's Maxwell 3D software, and the experimental results show that the design of permanent magnet can meet the requirements.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.