Abstract

The design and construction of inductive voltage dividers (IVDs) for use in impedance bridge applications in the 50 Hz to 1 MHz frequency range is described. Two dividers are described; one designed to operate from 50 Hz to 2 kHz and the other designed to operate from 2 kHz to 1 MHz. Preliminary comparison results of the new IVDs over the 100 Hz to 50 kHz frequency range are given. A new bridge to self-calibrate the IVDs is also described. The bridge is based on the straddling method and has been designed to characterize IVDs over the 50 Hz to 100 kHz frequency range.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call