Abstract

Following earlier work, a high temperature differential thermal calorimeter was constructed, with a design based on heat transfer considerations. Heat content and heat capacity measurements by the instrument showed excellent agreement with accepted literature values (±2%) in the case of quartz and alumina. Reasonable agreement was also obtained for the α->β quartz inversion enthalpy.