Abstract

Pulse oximeters may delay displaying the correct oxygen saturation during the onset of hypoxia. We investigated the desaturation response times of pulse oximeter sensors (forehead, ear and finger) during vasoconstriction due to mild hypothermia and vasodilation caused by glyceryl trinitrate. Ten healthy male volunteers were given three hypoxic challenges of 3 min duration under differing experimental conditions. Mild hypothermia increased the mean response time of finger oximeters from 130 to 215 s. Glyceryl trinitrate partly offset this effect by reducing the response time from 215 to 187 s. In contrast, the response times of the forehead and ear oximeters were unaffected by mild hypothermia, but the difference between head and finger oximeters was highly significant (p < 0.0001). The results suggest that the head oximeters provide a better monitoring site for pulse oximeters during mild hypothermia.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.