Abstract

The DEPFET pixel Bioscope system based on a DEPFET pixel matrix provides good spatial and energy resolution in real-time digital autoradiography. Due to its very low noise and a thin entrance window the online detection of tritium without vacuum or cooling is demonstrated for the first time. To achieve this milestone a p-channel junction field effect transistor on a fully depleted high ohmic silicon substrate (DEPFET) is used as unit cell for pixel detectors providing an excellent noise performance of 158 eV FWHM at 6 keV (12 e ENC) at room temperature (300 K). 64/spl times/64 DEPFET pixel matrices with square (50 /spl mu/m/spl times/50 /spl mu/m) or hexagonal (50 /spl mu/m/spl times/42 /spl mu/m) pixels have been developed and operated successfully. First measurements result in a homogeneous charge collection efficiency on the whole matrix area and a good linearity in the X-ray range from 5 keV to 60 keV. First images taken with the square and the hexagonal pixel matrices using an /sup 55/Fe-source as well as a 'proof of principle' experiment for online /sup 3/H-detection are presented and discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.