Abstract
High-pressure liquid chromatography (HPLC) performed in columns packed with aligned fiber can require less time and pressure drop than HPLC in traditional columns packed with particles, but resolution is limited by the uniformity of the packing. A model that accounts for the band spreading caused by fibers with distributed diameters predicts that the coefficient of variation (COV) of the chromatogram of an hexagonally closest packed aligned fiber column is about 1.2 times the COV of the diameter of the fibers constituting the column packing. To test the model, HPLC columns are prepared from bundles of aligned porous silica fibers tightly packed into steel tubes. The performance of the columns is in qualitative agreement with the theory. The dependence of band spreading on nonideal packing (i.e., packing that is not all of one geometric orientation as is hexagonal closest packing) can be a reason for differences between theory and experiment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.