Abstract

We report a new microcirculatory assessment device, the Braedius Cytocam, an Incident Dark Field (IDF) video microscope, and compare it with a precursor device utilising side stream dark field (SDF) imaging. Time matched measurements were made with both devices from the sublingual microcirculation of pigs subjected to traumatic injury and hemorrhagic shock at baseline and during a shock phase. Images were analysed for vessel density, microcirculatory flow and image quality. There were no differences in density or flow data recorded from the two devices at baseline [TVD IDF 14.2 ± 2.4/TVD SDF 13.2 ± 2.0, p 0.17] [MFI IDF 3 (2.8-3.0)/MFI SDF 3 (2.9-3.0), p 0.36] or during the shock state [TVD IDF 11.64 ± 3.3/TVD SDF 11.4 ± 4.0 p = 0.98] [MFI IDF 1.9 (0.6-2.7)/MFI SDF 1.7 (0.3-2.6) p 0.55]. Bland and Altman analysis showed no evidence of significant bias. Vessel contrast was significantly better with the IDF device for both capillaries [17.1 ± 3.9 (IDF) v 3.4 ± 3.6 (SDF), p = 0.0006] and venules [36.1 ± 11.4 (IDF) v 26.4 ± 7.1 (SDF) p 0.014] The Braedius Cytocam showed comparable vessel detection to a precursor device during both baseline and low flow (shock) states.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.