Abstract
Time-resolved X-ray absorption spectroscopy (TRXAS) beamline (BL2.2) was constructed to utilize a synchrotron radiation produced by a 1.2 T bending magnet from a 1.2 GeV storage ring at the Synchrotron Light Research Institute (SLRI), Thailand. The main optical component of the TRXAS beamline is an energy dispersive monochromator (EDM) which was donated by the University of Bonn, Germany. Using the energy-dispersive scheme with a position-sensitive detector, the beam size at the sample position can be small and have a fast detection speed in the XAS spectrum. At BL2.2, XAS measurement can be performed in the transmission mode using a 1024-pixel position-sensitive detector (PSD) and x-ray scintillator. Currently, the maximum measurable x-ray energy has been extended to 12 keV for the Pt L3-edge EXAS measurement. Due to the advantages of the energy-dispersive setup, this beamline has focused on in-situ XAS experiments. In this review, the researchers summarize the developments of experimental tools to characterize catalysts, Li-ion batteries, and thin film using TRXAS to determine their electronic properties and structure under applied conditions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.